Aria Gen2 Ambient Light Sensor (ALS)

A reference for researchers and developers working with ALS data from Aria Gen2 recordings. It covers the sensor itself, the on-device processing pipeline, what is stored in VRS, and how to read the data and factory calibration with projectaria_tools (PAT).


1. Sensor overview

Aria Gen2 uses the STMicroelectronics VD6281 RGB-IR-Clear-UV ambient light sensor. It is a 6-channel multispectral sensor with channels:

ChannelApproximate peakNotes
Red~620 nmColor channel
Green~545 nmColor channel
Blue~465 nmColor channel
UV~370 nmUltraviolet (UV-A region)
IR~850 nmNear-infrared
ClearbroadbandNo color filter, for illuminance

Each channel has its own optical filter and photodiode. The 6 channels are sampled simultaneously per frame.

Per-channel spectral response curves Sc(λ)S_c(\lambda) — the wavelength-dependent sensitivity of each filtered photodiode — are a property of the sensor hardware itself and are documented in the ST VD6281 datasheet. The datasheet is not openly distributed by ST; please contact ST (or your sales/FAE) to obtain a copy. Refer to the datasheet for the exact curve shapes.

Sample rate: nominally 10 Hz (configurable on-device).


2. Raw hardware output

For each frame, the sensor produces, for each channel cc independently:

QuantityTypeDescription
ADCc\text{ADC}_c24-bit (U16.8)Integrated photocurrent over the exposure window. Internally represented as 16 integer bits + 8 fractional bits.
gcg_c4-bit (1..15)Analog gain register code, mapping to one of 15 fixed gain values (see Normalization).
rer_e10-bitExposure time register code. The actual exposure time in microseconds is te=re×1600 μst_e = r_e \times 1600\ \mu s.

Each channel has its own programmable gain gcg_c — the sensor can use different gain settings on different channels to handle color sources where some channels would saturate faster than others.

The fundamental sensor equation (forward model):

ADCc=ChipGain(gc)I(λ)Sc(λ)dλte\text{ADC}_c = \text{ChipGain}(g_c) \cdot \int I(\lambda) \cdot S_c(\lambda)\, d\lambda \cdot t_e

where I(λ)I(\lambda) is the spectral irradiance of the incident light. The sensor integrates the spectrum × filter response over the exposure window and reports the result as a 24-bit count.


3. On-device processing pipeline

Before data is recorded, the device firmware applies the following processing to each frame:

3.1 Step 1 — Normalization (no calibration involved)

Nc=100,800 μsteADCcG[gc]\boxed{\quad N_c = \frac{100{,}800\ \mu s}{t_e} \cdot \frac{\text{ADC}_c}{G[g_c]} \quad}

where:

  • tet_e is the actual exposure time in µs
  • G[gc]G[g_c] is the analog gain value, looked up from the 15-entry table below

The constant 100,800 μs100{,}800\ \mu s is the manufacturer's "reference exposure". The result NcN_c is the equivalent signal as if the sample had been taken at that reference exposure with unity gain — making samples comparable across frames with different gain or exposure.

Gain lookup table G[g]G[g] (indices 1..15):

idxG[g]G[g]idxG[g]G[g]idxG[g]G[g]
166.67610.0111.67
250.077.14121.25
333.385.0131.0
425.093.33140.83
516.67102.5150.71

3.2 Step 2 — Linear flux mapping (uses calibration)

For the three flux-reporting channels (UV, IR, Clear), an affine transform maps the normalized signal to an irradiance value in W/m²:

Φk=αkNk+βk  [W/m2],k{UV,IR,Clear}\boxed{\quad \Phi_k = \alpha_k \cdot N_k + \beta_k\ \ [\text{W/m}^2], \qquad k \in \{\text{UV},\, \text{IR},\, \text{Clear}\} \quad}

The coefficients αk,βk\alpha_k, \beta_k come from the factory calibration (see Factory calibration: contents and usage).

3.3 Steps 3–5 — Lux and CCT (uses calibration)

A factory-calibrated 3×43{\times}4 matrix MM and 3-element offset vector o\mathbf{o} map four normalized channels (R, G, B, IR) into CIE XYZ tristimulus values:

[XYZ]=[m00m01m02m03m10m11m12m13m20m21m22m23]M[NRNGNBNI]+o\begin{bmatrix} X \\ Y \\ Z \end{bmatrix} = \underbrace{\begin{bmatrix} m_{00} & m_{01} & m_{02} & m_{03} \\ m_{10} & m_{11} & m_{12} & m_{13} \\ m_{20} & m_{21} & m_{22} & m_{23} \end{bmatrix}}_{M} \begin{bmatrix} N_R \\ N_G \\ N_B \\ N_I \end{bmatrix} + \mathbf{o}

CIE chromaticity coordinates:

x=XX+Y+Z,y=YX+Y+Zx = \frac{X}{X+Y+Z}, \qquad y = \frac{Y}{X+Y+Z}

Illuminance (in lux) — directly the Y tristimulus, since the matrix MM is fit to absorb the V(λ) photopic weighting:

L=max(Y,0)  [lux]\boxed{\quad L = \max(Y, 0)\ \ [\text{lux}] \quad}

Correlated color temperature (in Kelvin) — McCamy's 1992 cubic approximation:

n=x0.33200.1858yn = \frac{x - 0.3320}{0.1858 - y} Tcc=max ⁣(449n3+3525n2+6823.3n+5520.33, 0)  [K]\boxed{\quad T_{cc} = \max\!\Big(449\,n^3 + 3525\,n^2 + 6823.3\,n + 5520.33,\ 0\Big)\ \ [\text{K}] \quad}

This approximation is most accurate in the 2856 K – 6500 K range (incandescent through daylight).

A note on the matrix MM

It is a least-squares linear fit of (NR,NG,NB,NI)(X,Y,Z)(N_R, N_G, N_B, N_I) \to (X, Y, Z) trained against a set of standard calibration illuminants. It works well for typical broadband / "white" light sources. For narrow-band or unusual spectra (single-wavelength LEDs, lasers, line-emission lamps), the linear approximation breaks down because metameric spectra that differ from the calibration set are not captured. For such cases, you should work from the raw NcN_c values combined with the sensor's spectral response curves Sc(λ)S_c(\lambda) from the ST datasheet.

3.4 Auto-exposure

The firmware runs an on-device auto-exposure controller (AEC) that adjusts the exposure time on a per-sample basis to keep the brightest of the (R, G, B) channels in a healthy mid-range, well below the ADC saturation point. Channel gains are not adjusted by the AEC — they remain at fixed initial values for the lifetime of the recording.

Key consequences for users:

  • exposureTimeUs varies per frame in the range 1600 µs – 92800 µs. This is normal and expected.
  • All gain* fields are effectively constant across an entire recording: (gainRed, gainGreen, gainBlue, gainUv, gainIr, gainClear) = (1.0, 1.0, 1.0, 16.67, 1.0, 1.0).
  • Usable dynamic range is roughly 1 lux to ~20,000+ lux.
  • Per-sample coherence is preserved: the exposureTimeUs recorded for a given sample is exactly the exposure that was used to capture that sample. Auto-exposure changes only take effect on the next sample. The reverse-engineering formula in Recovering raw ADC counts is therefore exact.

If you need to detect frames that are saturated or at the AEC's exposure rails, see Detecting clipped / saturated samples.


4. Factory calibration: contents and usage

Each Aria Gen2 device is calibrated at the factory. The ALS-specific portion of the calibration is a JSON block with four entries:

JSON keyTypeUsed forCoefficient count
rgbarray of 12 or 15 floatsLux and CCT (matrix MM and offset o\mathbf{o})3×43 \times 4 matrix + 3 offsets = 15 (legacy: 12, IR column omitted)
uvarray of 2 floatsUV flux (W/m²)(αUV,βUV)(\alpha_{\text{UV}}, \beta_{\text{UV}})
irarray of 2 floatsIR flux (W/m²)(αIR,βIR)(\alpha_{\text{IR}}, \beta_{\text{IR}})
cleararray of 2 floatsClear flux (W/m²)(αClear,βClear)(\alpha_{\text{Clear}}, \beta_{\text{Clear}})

Layout of the rgb array (15-element form, row-major):

M_row_0 (X): [ m00, m01, m02, m03, o_X ]
M_row_1 (Y): [ m10, m11, m12, m13, o_Y ]
M_row_2 (Z): [ m20, m21, m22, m23, o_Z ]

The first 4 entries of each row are the matrix coefficients for (R, G, B, IR); the 5th entry is the offset for that row. In the 12-element legacy form there are only 4 columns (no IR column); the IR column is taken to be zero.

Where to find it: the calibration is embedded as a JSON string inside every Aria Gen2 VRS recording, in the factory_calibration field of each sensor stream's configuration record. See Reading factory calibration with PAT for how to read it.

On-device use: the firmware uses these coefficients to compute lux, CCT, and the three flux values before writing them to VRS. Researchers do not need to re-apply the calibration to get lux/CCT/flux. The coefficients are useful if you want to:

  • Recompute lux/CCT/flux from the recorded normalized channels (for verification),
  • Apply your own custom transform from the 6 channels to derived quantities,
  • Understand the relationship between raw channels and the cooked outputs.

5. What is stored in VRS

Each ALS sample in the VRS file contains the following fields:

FieldTypeUnitDescription
captureTimestampNsint64nsCapture timestamp (device monotonic clock)
redChannelNormalizedfloatunitlessNRN_R — normalized red channel
greenChannelNormalizedfloatunitlessNGN_G — normalized green channel
blueChannelNormalizedfloatunitlessNBN_B — normalized blue channel
uvChannelNormalizedfloatunitlessNUVN_{\text{UV}} — normalized UV channel
irChannelNormalizedfloatunitlessNIRN_{\text{IR}} — normalized IR channel
clearChannelNormalizedfloatunitlessNClearN_{\text{Clear}} — normalized clear channel
uvFluxWattPerSquareMeterfloatW/m²ΦUV\Phi_{\text{UV}} — UV irradiance
irFluxWattPerSquareMeterfloatW/m²ΦIR\Phi_{\text{IR}} — IR irradiance
clearFluxWattPerSquareMeterfloatW/m²ΦClear\Phi_{\text{Clear}} — broadband irradiance
gainRedfloatunitlessAnalog gain on red channel
gainGreenfloatunitlessAnalog gain on green channel
gainBluefloatunitlessAnalog gain on blue channel
gainUvfloatunitlessAnalog gain on UV channel
gainIrfloatunitlessAnalog gain on IR channel
gainClearfloatunitlessAnalog gain on clear channel
exposureTimeUsuint32µsExposure time used for this frame; varies per-sample in 1600–92800 due to on-device auto-exposure (see Auto-exposure)
cctfloatKCorrelated color temperature
luxfloatluxIlluminance
info

All six gain* fields are effectively constant across an Aria Gen2 recording — see Auto-exposure.

info

The PAT API also exposes three visible* fields (visibleChannelNormalized, visibleFluxWattPerSquareMeter, gainVisible) for cross-product schema compatibility. On Aria Gen2 these are always 0 and can be ignored.


6. Reading ALS data with projectaria_tools (PAT)

6.1 Basic iteration

from projectaria_tools.core import data_provider
from projectaria_tools.core.stream_id import StreamId

provider = data_provider.create_vrs_data_provider("recording.vrs")
als_sid = provider.get_stream_id_from_label("als")

# Total number of ALS samples
num_samples = provider.get_num_data(als_sid)
print(f"ALS samples: {num_samples}")

# Read one sample
sample = provider.get_als_data_by_index(als_sid, 0)
print(f"timestamp_ns = {sample.capture_timestamp_ns}")
print(f"lux = {sample.lux:.2f}")
print(f"cct = {sample.cct:.0f} K")
print(f"exposure_us = {sample.exposure_time_us}")
print(f"gain (RGB) = {sample.gain_red:.2f}, "
f"{sample.gain_green:.2f}, "
f"{sample.gain_blue:.2f}")
print(f"normalized = R:{sample.red_channel_normalized:.1f} "
f"G:{sample.green_channel_normalized:.1f} "
f"B:{sample.blue_channel_normalized:.1f} "
f"UV:{sample.uv_channel_normalized:.1f} "
f"IR:{sample.ir_channel_normalized:.1f} "
f"C:{sample.clear_channel_normalized:.1f}")
print(f"flux W/m² = UV:{sample.uv_flux_watt_per_square_meter:.4f} "
f"IR:{sample.ir_flux_watt_per_square_meter:.4f} "
f"C:{sample.clear_flux_watt_per_square_meter:.4f}")

6.2 Verify lux and CCT (sanity check)

Once you have the calibration matrix (see Reading factory calibration with PAT), you can recompute lux and CCT from the normalized channels and compare to the recorded values. They should agree to within float precision.

import numpy as np

def compute_lux_cct(N_R, N_G, N_B, N_I, M, offsets):
"""Apply factory calibration matrix to derive lux and CCT."""
XYZ = M @ np.array([N_R, N_G, N_B, N_I]) + offsets
X, Y, Z = XYZ
total = X + Y + Z
if total <= 0:
return 0.0, 0.0
x, y = X / total, Y / total
lux = max(Y, 0.0)
n = (x - 0.3320) / (0.1858 - y)
cct = max(((449 * n + 3525) * n + 6823.3) * n + 5520.33, 0.0)
return lux, cct

lux_recomputed, cct_recomputed = compute_lux_cct(
sample.red_channel_normalized, sample.green_channel_normalized,
sample.blue_channel_normalized, sample.ir_channel_normalized,
M, offsets)
assert abs(lux_recomputed - sample.lux) < 1e-3
assert abs(cct_recomputed - sample.cct) < 1e-1

6.3 Iterate all samples into a NumPy array

import numpy as np

n = provider.get_num_data(als_sid)
ts = np.zeros(n, dtype=np.int64)
lux = np.zeros(n, dtype=np.float32)
cct = np.zeros(n, dtype=np.float32)
normalized = np.zeros((n, 6), dtype=np.float32) # R, G, B, UV, IR, Clear

for i in range(n):
s = provider.get_als_data_by_index(als_sid, i)
ts[i] = s.capture_timestamp_ns
lux[i] = s.lux
cct[i] = s.cct
normalized[i] = [s.red_channel_normalized, s.green_channel_normalized,
s.blue_channel_normalized, s.uv_channel_normalized,
s.ir_channel_normalized, s.clear_channel_normalized]

7. Reading factory calibration with PAT

The factory calibration JSON is embedded into every sensor stream's configuration record. You can read it from any stream (image, IMU, ALS, etc.) — all streams in a recording carry an identical copy.

import json
from projectaria_tools.core import data_provider

provider = data_provider.create_vrs_data_provider("recording.vrs")

# Read from any sensor stream's configuration
rgb_sid = provider.get_stream_id_from_label("camera-rgb")
cfg = provider.get_image_configuration(rgb_sid)
factory_cal = json.loads(cfg.factory_calibration)

als_cal = factory_cal["als"]
rgb_coeffs = als_cal["rgb"] # 12 or 15 floats
uv_cal = als_cal["uv"] # [alpha_UV, beta_UV]
ir_cal = als_cal["ir"] # [alpha_IR, beta_IR]
clear_cal = als_cal["clear"] # [alpha_Clear, beta_Clear]

import numpy as np
if len(rgb_coeffs) == 15:
full = np.array(rgb_coeffs).reshape(3, 5)
M = full[:, :4] # 3x4 calibration matrix
offsets = full[:, 4] # 3-element offset vector
else: # legacy 12-coefficient format
M = np.zeros((3, 4))
M[:, :3] = np.array(rgb_coeffs).reshape(3, 4)[:, :3]
M[:, 3] = 0.0 # IR column = 0
offsets = np.array(rgb_coeffs).reshape(3, 4)[:, 3]

print(f"3x4 matrix M:\n{M}")
print(f"offsets: {offsets}")
print(f"UV cal: alpha = {uv_cal[0]}, beta = {uv_cal[1]}")
print(f"IR cal: alpha = {ir_cal[0]}, beta = {ir_cal[1]}")
print(f"Clear cal: alpha = {clear_cal[0]}, beta = {clear_cal[1]}")

You can then use the calibration to verify flux values:

def expect_close(name, expected, actual, tol=1e-5):
assert abs(expected - actual) < tol, f"{name}: {expected} vs {actual}"

alpha_U, beta_U = uv_cal
alpha_I, beta_I = ir_cal
alpha_C, beta_C = clear_cal

expect_close("UV flux",
alpha_U * sample.uv_channel_normalized + beta_U,
sample.uv_flux_watt_per_square_meter)
expect_close("IR flux",
alpha_I * sample.ir_channel_normalized + beta_I,
sample.ir_flux_watt_per_square_meter)
expect_close("Clear flux",
alpha_C * sample.clear_channel_normalized + beta_C,
sample.clear_flux_watt_per_square_meter)

8. (Optional) Recovering raw ADC counts

Although VRS stores normalized values rather than raw ADC counts, the raw counts can be reconstructed by inverting the normalization formula in Normalization:

ADCc=NcG[gc]te100,800\boxed{\quad \text{ADC}_c = \frac{N_c \cdot G[g_c] \cdot t_e}{100{,}800} \quad}
LUT_GAIN = [66.67, 50.0, 33.3, 25.0, 16.67, 10.0, 7.14, 5.0,
3.33, 2.5, 1.67, 1.25, 1.0, 0.83, 0.71]

def gain_to_index(g):
"""Map the float gain stored in VRS back to the 1..15 index."""
return min(range(15), key=lambda i: abs(LUT_GAIN[i] - g)) + 1

# Internal-precision gain values (8.8 fixed-point used by the on-device driver)
LUT_GAIN_EXACT_HEX = [0x42AB, 0x3200, 0x2154, 0x1900, 0x10AB, 0x0A00,
0x0723, 0x0500, 0x0354, 0x0280, 0x01AB, 0x0140,
0x0100, 0x00D4, 0x00B5]

def exact_gain(idx):
return LUT_GAIN_EXACT_HEX[idx - 1] / 256.0

def recover_raw_adc(N_c, gain_float, exposure_us):
"""Returns the 24-bit U16.8 raw ADC representation used by the driver.
Approximate 16-bit ADC count: int(result) >> 8.
"""
idx = gain_to_index(gain_float)
G = exact_gain(idx)
return N_c * G * exposure_us / 100_800.0

t_e = sample.exposure_time_us
adc_red = recover_raw_adc(sample.red_channel_normalized, sample.gain_red, t_e)
adc_green = recover_raw_adc(sample.green_channel_normalized, sample.gain_green, t_e)
adc_blue = recover_raw_adc(sample.blue_channel_normalized, sample.gain_blue, t_e)
adc_uv = recover_raw_adc(sample.uv_channel_normalized, sample.gain_uv, t_e)
adc_ir = recover_raw_adc(sample.ir_channel_normalized, sample.gain_ir, t_e)
adc_clear = recover_raw_adc(sample.clear_channel_normalized, sample.gain_clear, t_e)

Precision note: the on-device firmware casts the normalized value from a 64-bit fixed-point representation to float32 before recording. This is the only precision-loss step in the entire pipeline:

  • For typical indoor illuminance, reconstructed ADC values are accurate to integer level.
  • For bright outdoor or near-sun readings (normalized values approaching 5.8×1065.8 \times 10^6), up to ~14 bits of fractional precision are lost.

If you require bit-exact raw counts, please contact the Aria team — that would require a firmware change to also emit the underlying fixed-point value.

8.1 Detecting clipped / saturated samples

The device does not record an explicit saturation flag per sample. If your analysis is sensitive to channel saturation (e.g. characterizing bright outdoor scenes), it is recommended to derive one yourself from the reconstructed raw ADC and the sensor's documented saturation rate of approximately 145.636 ADC counts / µs (with an 80% safety margin applied by the on-device AEC):

SAT_RATE_ADC_PER_US = 145.636
SAFETY_MARGIN = 0.80

def sample_quality(sample):
t_e = sample.exposure_time_us
sat_threshold = SAT_RATE_ADC_PER_US * t_e * SAFETY_MARGIN
raw_R = recover_raw_adc(sample.red_channel_normalized, sample.gain_red, t_e)
raw_G = recover_raw_adc(sample.green_channel_normalized, sample.gain_green, t_e)
raw_B = recover_raw_adc(sample.blue_channel_normalized, sample.gain_blue, t_e)
raw_I = recover_raw_adc(sample.ir_channel_normalized, sample.gain_ir, t_e)
return {
'rgb_saturated': max(raw_R, raw_G, raw_B) > sat_threshold,
'ir_saturated': raw_I > sat_threshold,
'aec_at_min_exp': t_e <= 1600, # scene may be brighter than the device can resolve
'aec_at_max_exp': t_e >= 92800, # scene may be darker than the device can resolve
}

Use this to filter out frames at the edges of the sensor's dynamic range before downstream analysis.


9. Live streaming

Live streaming of ALS data is not currently supported for external users. If real-time ALS streaming is required for your work, please contact us at AriaOps@meta.com to discuss your use case.

Version requirements

projectaria_tools ≥ 2.1.4 is required for the gain* and exposureTimeUs fields to read back correctly. Earlier versions had a schema-type mismatch that caused these fields to silently return 0. Please upgrade if your installed version is older.


10. References

Sensor datasheet

  • STMicroelectronics VD6281 (not openly distributed by ST; contact ST or your sales/FAE to request a copy)

projectaria_tools (Apache 2.0, open source)

Standards and references

  • CIE 1931 colorimetry (X, Y, Z tristimulus and chromaticity)
  • McCamy, C. S. (1992). "Correlated color temperature as an explicit function of chromaticity coordinates." Color Research & Application, 17(2), 142–144.